Determination of Mefenamic Acid using 8-hydroxy quinoline as a Precipitating Agent and Low Pressure Mercury Lamp (184.9 & 253.7 nm) as a Source of Irradiation using of ISNAG Continue Flow Fluorimeter

Hussein Fares Abd-Alrazack

Abstract

A new, simple, sensitive and fast method for the determination of Mefenamic acid in pure form and drugs (tablets) by continuous flow feed analysis via measurement of diverged beam of light by ISNAG_-fluorimeter (homemade instrument). The method based on the reaction of the Mefenamic acid with 8-HQ to form yellow precipitate. Optimum parameters have been studied to increase the sensitivity for the developed method. The linear dynamic range for the instrument response versus Mefenamic acid concentration was 0.005-7 mmol/L while the L.O.D was 0.216 µg/sample from the step wise dilution for the minimum concentration of lowest concentration in the linear dynamic range of the calibration graph. The correlation coefficient (r) was 0.9956 while percentage linearity (R2%) was 99.12%. RSD % for the repeatability (n=8) was lower than 0.2% for the determination of Mefenamic acid, with concentration of 2, 5 mmol/L respectively. The developed method was applied successfully for the determination of Mefenamic acid in pharmaceutical tablets. A comparison was made between the newly developed method with the classical method (UV-Vis spectrophotometry at wavelength 285nm, and Turbidemtric method) of analysis using the standard addition method via the use of paired t-test. It shows that there was no significant difference between the quoted values of each individual company with calculated t-value at 95% confidence interval from developed method.

Keywords: Mefenamic acid, Flow injection analysis, Homemade instrument.

Full Text:

PDF

Refbacks

  • There are currently no refbacks.